CFT-003058 (ID15A High resolution sample stage)

Opening date

March 5, 2021

Closing date

April 2, 2021

Type

Forthcoming tender

Research Facility

ESRF

This call for tender concerns the design, manufacture and supply of a high-resolution sample stage assembly for high-speed high spatial resolution diffraction imaging/tomography for the ID15A-EH3 experimental station. During the data acquisition, which can last several hours, the sample is simultaneously continuously rotated and translated across the beam. A sample positioning accuracy of ~100nm over a translation range of ~1 mm and a rotation of 360 degree is required with samples weighing ~200 g. The sample stage shall consist of: an air bearing rotation stage (Rz) with maximum rotation speed of ~5 rpm; an air bearing translation stage (Ty); a vertical translation stage (Tz); a tilt stage (Ry); electrical slip rings for passing current through the rotary stage to the sample cell. A 4 small axis translation (Tx2 & Ty2) and tilt stage (Rx2 & Ry2) for sample pre-alignment should be mounted on top of the rotation stage. The motion of the Rz and Ty stages must be synchronized.